Planck-LFI: Design and Performance of the 4 Kelvin Reference Load Unit
The LFI radiometers use a pseudo-correlation design where the signal from the
sky is continuously compared with a stable reference signal, provided by a
cryogenic reference load system. The reference unit is composed by small
pyramidal horns, one for each radiometer, 22 in total, facing small absorbing
targets, made of a commercial resin ECCOSORB CR (TM), cooled to approximately
4.5 K. Horns and targets are separated by a small gap to allow thermal
decoupling. Target and horn design is optimized for each of the LFI bands,
centered at 70, 44 and 30 GHz. Pyramidal horns are either machined inside the
radiometer 20K module or connected via external electro-formed bended
waveguides. The requirement of high stability of the reference signal imposed a
careful design for the radiometric and thermal properties of the loads.
Materials used for the manufacturing have been characterized for thermal, RF
and mechanical properties. We describe in this paper the design and the
performance of the reference system.Comment: This is an author-created, un-copyedited version of an article
accepted for publication in JINST. IOP Publishing Ltd is not responsible for
any errors or omissions in this version of the manuscript or any version
derived from it. The definitive publisher authenticated version is available
online at [10.1088/1748-0221/4/12/T12006]. 14 pages, 34 figure